Description of EMC
B-1
Appendix B
Guidance and Manufacturers Declaration
Electromagnetic Emissions for All Equipment and Systems
To provide reasonable protection, this product complies with emissions limits for a Group
1, Class A ISM equipment Medical Devices Directive as stated in EN 60601-1-2.
Warning 1: The use of accessories, sensors, and/ or cables not supplied by SIUI
may lead to increase of electromagnetic radiation or decrease of electromagnetic
immunity, consequently resulting in operation failure.
Warning 2: The distance between any part of SR-8210 (including cables) and
portable RF communication equipment shall not be less than 30cm, otherwise it
may result in declined device performance.
Warning 3: Precaution to be taken to prevent adverse events to the PATIENT and
OPERATOR due to ELECTROMAGNETIC DISTURBANCES is to ensure that the
main unit is grounded reliably.
Note: The EMISSIONS characteristics of SR-8210 make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment SR-8210 might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as
relocating or re-orienting the equipment.