General
1-3
1.8 Impact to Environment and Energy
a) Power-saving technology is used in the device. For example, instead of
a conventional halogen lamp, a high-bright LED is used in the collimator, with the
power only a few tenth of the halogen lamp.
b) The device shall be recycled properly at the end of its shelf life in accordance with
relevant national safety and environmental standards & regulations, so as to
reduce the risks arising to the lowest level.
1.9 Classification
1.9.1 Classification by anti-shock type
The product is a Class I internal power supply device.
1.9.2 Classification by anti-shock degree
The product is a common device of Type B applied part.
1.9.3 Classification by IP rating
The product IP rating is IPX0.
1.9.4 Classification by safe use in the case of flammable anesthetic gas mixed with
air or flammable anesthetic gas mixed with oxygen or nitrous oxide
DO NOT USE THE DEVICE in the presence of flammable anesthetic gas mixed with air or
with oxygen or nitrous oxide.
1.9.5 Classification by operating mode
Continuous operation, intermittent loading.
1.10 Safety
a) The device does not apply to mammography and dental radiography.
b) It is prohibited for the same patient (especially children) to receive high-frequency
imaging of the device for a long period.
c) Pregnant women undergoing X-ray exams may result in fetal injury. Clinicians’
decision shall be made by weighing the condition of patients.