Use and Operation
6-21
Standby state. You can change the time for P6 “Standby
time” and P21 ”Timing setting from shutdown to poweroff” in
Engineering Mode in 6.1.3 "Engineering Mode Function and
Operation".
d) There is an icon in the upper right of the portable X-ray
machine for power display. When the power is low, recharge
is required in time. Make sure that the power switch on the
back of the portable X-ray machine is OFF.
e) To protect the X-ray machine, the intelligent temperature
control system will impose some restrictions during
exposure. If consecutive exposures are performed when
mAs≥30, the time interval after each exposure will increase.
At this time, stop the exposure operation and take a break
for a few minutes, or perform an exposure with mAs <30
during consecutive exposure, which can decrease the
waiting time interval after exposure.
f) To protect the X-
ray machine, the intelligent protection
system will impose some restrictions during exposure. If the
remaining
battery level is displayed in one block, the
available range of mAs will be decreased.