Safety Protection Device and Troubleshooting
7-3
Code
Failure Description and Cause
Analysis
Troubleshooting
E09
Abnormal parameters of tube
voltage self-check in standby state
Press Fn to clear the code. If the error
still exists, turn off the system and contact
the manufacturer for repair.
E10
Ab
normal parameters of tube
current self-check in standby state
Press Fn to clear the code. If the error
still exists, turn off the system and contact
the manufacturer for repair.
E16
No tube voltage output in
exposure.
Press Fn to clear the code. Perform
exposure after 10 minutes’ system break.
If the failure reproduces, check whether
battery power is too low.
HBA
Press Fn to clear the code. If the error
still exists, turn off the system and contact
the manufacturer for repair.
LBA
Press Fn to clear the code. If the error
still exists, recharge the battery.
HET
Exposure interval time limit
It would be automatically clear when the
exposure interval was over.
If the
exposure interval is not finished, the
device will still re
shutting down and restarting.
EL0
Fail to export the user exposure
parameters to the USB disk during
the program update.
The error code will be cleared
automatically after flashing for 1 second.
EL1
Fail to export the APR parameters
t
o the USB disk during the
program update.
The error code will be cleared
automatically after flashing for 1 second.
EL2
Fail to export the filament value to
the USB disk during the program
update.
The error code will be cleared
automatically after flashing for 1 second.
EL3
Fail to export the system state
parameters to the USB disk during
the program update.
The error code will be cleared
automatically after flashing for 1 second.
EL4
Fail to export the error reporting
function mark to the USB disk
during the program update.
The error code will be cleared
automatically after flashing for 1 second.