Description of EMC
B-3
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test IEC 60601 test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF
electromagnetic
field immunity
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Proximity Fields
from RF Wireless
Communications
Equipment Test
IEC 61000-4-3
Test Frequency
(MHz)
Modulation Immunity Level (V/m)
385 **Pulse Modulation:
18Hz
27
450 *FM
± 5kHz deviation: 1kHz
sine
28
710
745
**Pulse Modulation:
217Hz
9
810
870
**Pulse Modulation:
18Hz
28
1720
1845
**Pulse Modulation:
217Hz
28
2450 **Pulse Modulation:
217Hz
28
5240
5500
5785
**Pulse Modulation:
217Hz
9
Conducted RF
IEC 61000-4-6
3 V RMS outside the ISM band 0.15 MHz to 80 MHz
6 V RMS in the ISM band 0.15 MHz to 80 MHz
80 % AM at 1 kHz
Power frequency
magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
NOTE: U
T