38
Accuracy: (0.1 + 0.05 sec)
Range: 0 – 100 mΩ
Resolution: 1 mΩ
Accuracy: (2 of setting + 2 mΩ)
Input: Test, Reset, Interlock, Recall Memory 1 - 6
Output: Pass, Fail, Test-in-Process
6 memories, 6 steps per memory
All steps are linkable
Single step mode
Key Lock capability to avoid unauthorized access to all test parameters.
Memory Lock capability to avoid unauthorized access to memory locations.
Detachable 6 ft (1.8 m) power cable terminated in a three-prong grounding
plug.
6 ft (1.8 m) high voltage and return leads (2) with clips. An optional remote
receptacle box may be purchased for testing items terminated with a line
cord. International receptacles are also available.
Tilt up front feet.
Dimensions: (WxHxD) 16.9 x 5.2 x 15.7 inches (430 x 133 x 400 mm)
Weight: 54 lbs (24.5 Kg)
Operating Temperature : 32 - 104F (0 - 40C)
Relative Humidity: 0 to 80%
Traceable to National Institute of Standards and Technology (NIST).
Calibration controlled by software. Adjustments are made through front
panel keypad in a restricted access calibration mode. Calibration
information stored in non-volatile memory.
Why use the term “Counts”?
Slaughter publishes some specifications using COUNTS which allows us to provide a better indication of the
tester’s capabilities across measurement ranges. A COUNT refers to the lowest resolution of the display for a
given measurement range. For example, if the resolution for voltage is 1V then 2 counts = 2V.