EasyManua.ls Logo

SolarMax MaxWeb XPN - Page 55

SolarMax MaxWeb XPN
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
55
en
Fig. 52: Conguration of external alarm transmitters
Alarm 1 / 2: Tick this option if you have connected external alarm sensors the data
of which is to be logged.
Input:
Select the terminals to which the external alarm sensor is connected
in this menu. For a detailed overview of the assignment and designa-
tion of terminals, see Section 8.5.
Delay: In this menu, enter the number of seconds by which the signal must
be delayed for an alarm to be triggered.
If a MaxDisplay is connected, mark the “Active” checkbox in the “MaxDisplay” menu.
Fig. 53: Activation of the MaxDisplay
The operator can connect devices by using the relay outputs of the MaxWeb XPN which
can be operated with 30V/1A. These devices will be connected to the upper terminal
clamps on the rear side of the MaxWeb XPN (see Section 8.5.2).

Related product manuals