Chapter 6: Technical Information
Methodology of Reagent
When glucose reacts with the reagents on the test strips, an electrical current is produced,
which is proportional to the glucose concentration in the blood sample.
The glucose concentration is calculated by the meter type and the amount of electrical
current measured.
Limitations
1. Do not use for neonatal blood glucose testing.
2. Hematocrits in the range of 32 to 56 percent do not affect the blood glucose results. If
you do not know your hematocrit level, consult your healthcare professional.
3. Hemoglobin levels of 500 mg/dL or below will not interfere with blood glucose test
results.
4. Cholesterol levels up to 500 mg/dL and Triglycerides up to 1000 mg/dL have been shown
not to affect glucose results.
5. Medications: Interference was observed for therapeutic levels of L-Dopa. No interference
was shown for uric acid, acetaminophen, ascorbic acid and ibuprofen at normal
therapeutic levels; however, higher concentrations in blood may cause incorrect results:
Uric acid: >10.9 mg/dL;
Acetaminophen: >6.2 mg/dL;
Ascorbic acid: >4.5 mg/dL; or
Ibuprofen: >37.5 mg/dL.
6. The test strips may be used at altitudes up to 5,280 feet (1,609m) without effecting test
results.
7. If you are suffering from severe dehydration you should have your blood glucose levels