EasyManua.ls Logo

Solus V2 User Manual

Solus V2
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Safety Information
Please use this device only for the intended uses described in this User Guide.
Before using this system to test your blood glucose, please read instructions thoroughly. Do a
quality check on the system following the instructions and consult with your healthcare
professionals about questions or problems.
Be aware of the safety of young children or handicapped persons near you when you
conduct a glucose test using this system.
SOLUS test strips can only be used with the Solus V2 meter.
Please keep the test strip bottle away from children. The test strips and bottle cap can
present a choking hazard.
Please be cautious when removing the lancet. Always place the protective cap back on the
exposed lancing device.
Never try to disassemble the meter in any circumstance. If your meter is not working
properly, or if you need technical support, please contact our Technical Support Team at (877)
592-3922.
Rev. Date: 06/2012 6131SL-M01A2
Question and Answer IconNeed help?

Do you have a question about the Solus V2 and is the answer not in the manual?

Solus V2 Specifications

General IconGeneral
Blood Sample Size0.5 μL
Dimensions96 x 61 x 24 mm (3.8 x 2.4 x 0.9 in)
Weight72 g (2.5 oz) with battery
Operating Humidity10 - 90% RH
BatteryTwo 1.5V AAA alkaline batteries
Measuring Range20 - 600 mg/dL
Operating Temperature10 - 40°C (50 - 104°F)

Summary

Safety Information

Introduction

Intended Use/ Indications for Use

Details the purpose, intended users, and limitations of the Solus V2 Blood Glucose Management System.

Understanding Your New System

SOLUS Audible Blood Glucose Management System

Lists the components included in the Solus V2 Blood Glucose Management System package.

Installing/Replacing the Batteries

Provides step-by-step instructions for safely removing and replacing the meter's AAA batteries.

Meter Setup

Outlines the customizable features for personalizing the meter settings like volume, language, and date/time.

Setting the Volume Level and Language Selection

Explains how to adjust the meter's audio volume and switch between English and Spanish languages.

Setting the Meter

Details the process of setting the meter's date, time, and alarms.

LCD Screen Overview

Explains the meaning of various symbols and indicators displayed on the meter's LCD screen for user reference.

Setting the Alarm Reminder

Instructions on how to enable, disable, or set a specific time for the alarm reminder function.

Setting the Measurement Unit

Explains how to select between mg/dL and mmol/L measurement units.

Deleting the Memory

Provides instructions for deleting stored test results from the meter's memory.

Performing a Control Test

Why and When to Perform a Control Test

Explains the importance, purpose, and recommended frequency of performing control solution tests.

About Control Solution Tests

Details the proper preparation steps and procedures for conducting a control solution test accurately.

Performing a Control Test

Step-by-step guide on how to execute a control solution test using the Solus V2 meter.

Testing Your Blood Sugar

Preparing the Solus V2 Lancing Device

Instructions on how to safely insert, adjust, and prepare the lancing device for blood sampling.

Performing a Blood Glucose Test

Step-by-step guide on how to perform a blood glucose test using the Solus V2 meter and test strips.

Alternate Site Testing (AST) Information

Information on Alternate Site Testing (AST), including its definition, advantages, and usage conditions.

Performing an Alternate Site Test (AST)

Step-by-step guide on how to perform a blood glucose test at an alternate site on the body.

Highlighting Results as Before and After Meals

Instructions on how to flag test results as 'before meal' or 'after meal' for better tracking.

Understanding Your Test Results

Explains how to interpret blood glucose readings and provides normal ranges for context.

Symptoms of High or Low Blood Sugar

Lists common symptoms associated with hyperglycemia (high sugar) and hypoglycemia (low sugar).

Unusual Test Results

Provides troubleshooting steps to follow when blood sugar results do not match the user's condition.

Recalling the Memory and Downloading Results

Recalling the Memory

Instructions on how to access and navigate through stored test results from the meter's memory.

Viewing the Averages

Details how to view average blood glucose readings over different specified time periods (7, 14, 28, 60, 90 days).

Displaying Averages in Sequential Order

Explains how to view 'before-meal' and 'after-meal' averages sequentially.

Downloading Test Results with MyMeterLink

Guide to transferring stored test results from the meter to a computer via the MyMeterLink system.

Maintenance and Troubleshooting

Storing and Handling Your Meter

Guidelines for the proper storage and handling of the Solus V2 meter and its test strips.

Cleaning Your Meter

Instructions on how to safely and effectively clean the surface of the Solus V2 Glucose Meter.

Interpreting Error Messages

Explains common error messages, their possible causes, and recommended solutions for meter issues.

Troubleshooting Your Meter

Covers general operational problems and provides solutions for issues not covered by specific error codes.

Technical Information

Accessing Customer Service

Contact information and support details for obtaining assistance with the Solus V2 Glucose Management System.

Methodology of Reagent

Explains how the reagents on the test strips react with blood to produce an electrical current for measurement.

Limitations

Lists factors, such as specific medications or conditions, that may affect the meter's test results.

Product Specifications

Details the technical specifications of the Solus V2 Glucose Meter, including range, time, and dimensions.

Warranty Information

Outlines the warranty period for the Solus V2 Glucose Meter and provides contact information for claims.