tested by a healthcare professional in a hospital or lab setting. Do not test with a
home-use blood glucose device when suffering from severe dehydration.
8. Test results below 70 mg/dL indicate low blood glucose (hypoglycemia). Test results
greater than 240 mg/dL indicate high blood glucose (hyperglycemia). If you get results
below 70 mg/dL or above 240 mg/dL, repeat the test, and if the results are still below 70
mg/dL or above 240 mg/dL, please consult your healthcare professional immediately.
2
9. Inaccurate results may occur in severely hypotensive individuals or patients in shock.
These results may occur for individuals experiencing a hyperglycemic-hyperosmolar state,
with or without ketosis. Critically ill patients should not be tested with a blood glucose
meter.
Product Specifications
Test: Glucose
Sample: Whole Blood
Principle of the test method: Amperometric, glucose oxidase.
Test sites: Fingertip, Forearm
Measuring Range: 20–600 mg/dL (1.1–33.3 mmol/L)
Measuring Time: 6 seconds
Code required: No code device
Batteries: ALKALINE 24A LR03 AAA 1.5V X 2
Operating Range: 10–40
o
C (50–104
o
F)
Humidity: 20–80 % RH
Width: 55 mm.
Length: 100 mm.
Thickness: 18 mm.
Weight: 75 g.
Memory: 500 test results with date and time.
Power: Automatic shut-off after five (5) minutes when a strip is inserted, or one (1) minute if
no strip is inserted.
NDC #: 8611-501001
The device has been certified to meet the following standard:
98/79/EC, EN 60601-1 + EN 60601-1-1
ISO 15197