EasyManuals Logo

Sonoscape S6 Service Manual

Sonoscape S6
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
S6 Portable Digital Color Doppler Ultrasound System
Service Manual
7.4.2 Non-kernel Level Software Update Procedures
1. Plug the USB disk drive with update files into the USB port of the
system.
2. Boot up the system, wait until the EXAM screen appears.
3. Press these keys successively: Shift, E and Menu. The system will
require a password to continue:
4. Enter the correct password (the default password is HH77 for year
2009), the update confirmation dialogue is shown as below.
5. Select [GO] and press SET key to continue (Press Exam key will can-
cel the update process).
6. The system will show this message: “Please wait. . . ”.
7. Wait until the message “Shut Down System” is displayed on the screen.
Press Exam key, remove the USB disk and restart the system to com-
plete the update.
Note:
1. The password used in the update process is provided to authorized
service/maintenance personals by SonoScape. The password is NOT
case sensitive.
2. If the message, “Check the USB”, pops up, reinsert the USB disk or
use the other USB port.
3. Do not plug in any removable disks into it while the system is starting
up to avoid any interruption.
4. The above update process will not change the local cine files or the
patients data.
P/N: 4720-0034-01A
7-4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonoscape S6 and is the answer not in the manual?

Sonoscape S6 Specifications

General IconGeneral
BrandSonoscape
ModelS6
CategoryMedical Equipment
LanguageEnglish

Related product manuals