S6 Portable Digital Color Doppler Ultrasound System
Service Manual
7.5.4 Activate the µ-SCAN function
Attention!
In order to use the µ-SCAN function, any external USB disk
drives should be removed before the system startup.
a. Insert the µ-SCAN dongle into the designated USB connector (refer to
section 5.3 to connect).
b. Copy the Keypass file from PC to the root directory of a USB disk.
c. Remove any removable USB drives if any, start up the S6 system.
d. Plug in the USB disk drive with Keypass file when the EXAM screen
appears.
e. Enter engineer mode by pressing these keys successively: Shift, E and
Menu, and then type in the monthly engineer password.
f. Move the cursor above IN button, and press SET key to write in the
Keypass file. After successfully writing in, press EXAM key to exit.
g. Re-entering engineer mode, select Engineer Config. Press ‘I’ button to
activate Keypass. The successful activation will lead to the following
screen. (If the activation fails, check whether the Keypass file is con-
sistent with the µ-SCAN dongle and located at the root directory of the
USB disk.)
h. Remove any removable USB drives, restart the system. Enter a diag-
nostic mode and press MENU key, the µ-SCAN functionality should be
activated.
Note: If the µ-SCAN function has not been enabled up till now, ensure that
the µ-SCAN dongle is correctly plugged in and repeat procedures from step
d above. In order to use the µ-SCAN function, no external USB drives
should be connected during system startup.
P/N: 4720-0034-01A
7-10