| Troubleshooting | 43
Troubleshooting
Troubleshooting FX
This section provides troubleshooting instructions for FX series.
Symptom Possible Cause Solution
Camera does not start, LED is
blinking.
Boot loop Remove camera power cord from the
wall socket, wait a few minutes, then
reconnect the power.
No signal to data acquisition PC. Lens is covered. Remove the lens cap.
Non-standard lens aperture is too
small
Check if the camera lens is standard.
If not
• Check if the lens aperture can be
changed
• Check that the aperture is not too
small
Image quality is poor. Insufficient illumination.
• Add lighting.
• Lengthen the integration time.
Camera shuts down unexpectedly. Operating temperature is too high. Disconnect the power cord and let
the camera cool down before trying
to power it up again. Add cooling if
the temperature rises repeatedly.
Data is not stored to PC. Save folder is not set. Define the save folder in LUMO.
All rights reserved - Specim, Spectral Imaging Oy Ltd.