| Technical Information | 50
½ sine 40 g, 11 ms, with 3 shocks per axis per direction.
Temperature test
The temperature test has been performed in temperature cycle, run for 2 weeks. The test cycle was as follows:
Ramp to 5°C (30 min)
Steady 5°C for 240 min
Ramp to 40°C (30 min)
Steady 40°C for 240 min
Ramp to 22°C (30 min)
All rights reserved - Specim, Spectral Imaging Oy Ltd.