Cirrus SP101013.111 – 10 October 2012
MKS Instruments UK Ltd., Spectra Products
RGA Comms
An RJ-45 socket is used for connection to the host PC.
A Cat5 STP cable MUST be used for connection to this instrument in order to comply with the immunity
requirements of the EMC directive.
See Page 4 for maximum cable lengths.
Power
The mains power switch and fused IEC socket.
Analog I/O
The Analog I/O connector is a 15-way D-Type socket.
See Page 4 for maximum cable lengths.
The Analog I/O port provides:
One analog output 0 to +10V
Four, quasi-differential analog inputs, 0 to ±10V, with a maximum voltage on the return of
±0.5V
±15V power outputs both fused at 100mA, fuses are not self resetting.
Analog Connector pin assignment.
Analog input 1 (Reserved)
Note: The total power consumption on each rail (+5V and ±15V) for both the Analog and Digital I/O
ports must not exceed 100mA.
* Analog input 1 and Analog input 1 return are reserved for use with the internal cold cathode gauge.