Setting “All Pumps Off” on the Master will change the Remote to
All Pumps Off and flash the display. This can then be overridden
at the Remote with the APO button. Pumps not controlled by the
Master may continue to operate on the Remote.
Input #3 [Rem]
Pumps: _ B _ D
IF Sensor: WET
THEN Pump: OFF
D:Settings
Mode: DOSE [Rem]
Each: 10 ml
Every: 20 mins
Change Input #3 from [Loc] to [Rem], enable
the Remote’s sensor #3 and attach Pumps B
and D to it. When the Remote sensor is wet,
the pumps will turn off.
Master controls the Remote’s Pump D
in DOSE Mode, delivers 10 ml every 20
minutes. [Rem] is displayed whenever
Remote’s Pump D Mode is set to “ON”.
Some Examples Using Remote Mode:
Master LM4 Screen:
Master LM4 Screen:
D:Settings
Mode: ON
Under Master
Control
Remote’s Pump D is enabled by the
Master LM4 and is running under
Master control.
Remote LM4 Screen:
D:Settings
Mode: Disabled
Under Master
Control
Remote Pump D is disabled by the
Master LM4 and is NOT running.
Remote LM4 Screen:
B:Settings
Mode: DAILY
Total: 12.00 L
Cycles: 24/day
Remote Pump B is also running in its own
DAILY Mode, 12 Liters per day / 500 mL
per cycle.
Remote LM4 Screen: