SS1000 Operator’s Manual
3
01-92-4 A .ver 97020-000942
3. After initialization, the LCD displays the Normal Mode screen with
four lines (the third of which is blank for single-channel units).
4. Enable Peak Tracking following the procedure outlined in “To
change parameters in Mode 2 on page 3-6.
5. Continuous updates of the measurement parameters displaying on
the LCD indicate that the analyzer is operating normally.
6. Allow approximately three additional minutes for the analyzer to
stabilize before recording measurements.
Powering Down the Analyzer
It may be necessary to power down the analyzer for problem solving or
maintenance reasons.
To power down the analyzer:
1. Power down the analyzer by turning off the power switch on the back
of the analyzer.
2. If the analyzer is going to be shut down for a long period of time,
turn off sample flow to the analyzer by closing the sample probe
shut-off valve at the sample supply tap.
3. Allow pressure in the field-pressure reducing regulator to dissipate
until only a low residual pressure is indicated on the pressure gauge
at the field station.
4. C
lose the field-pressure reducing regulator (adjustment knob turned
fully counterclockwise).
Operating the Analyzer from the Keypad
The keypads enable the operator to modify measurement units, adjust
operational parameters, and perform diagnostics. During normal operation, the
LCDs continuously display the measured components’ concentrations, sample
cell temperatures, and sample cell pressures.
<NORMAL MODE>
H2O: 1.0256lb/mmscf
P: 954.4mb T: 76.1F