SS1000 Operator’s Manual
3
– 01-92-4 A .ver 97020-000946
Mode TEST: (System Test)
Provides basic diagnostic test results indicating that the laser power,
pressure/temperature sensors, and the infrared spectrum that the system
records for analysis are all nominal. Press the # key followed by the TEST key.
The LCD displays the system test data.
Changing Measurement and Control Parameters
In Mode 2, all of the pertinent measurement and control parameters can be
viewed and changed. Refer to Table 3–1 for a list of parameters and value
ranges.
To change parameters in Mode 2:
1. Press the # key followed by the 2 key. The LCD prompts for a
numeric password.
2. Enter the user password (3142) on the keypad, then press the * key
to enter the number.
3. Starting with the first parameter, enter a new value and/or press the
* key to store the value and cycle to the next parameter.
+
#
TEST
<SYSTEM TEST>
Laser Power: OK
PT Sen: OK Trak: OK
Spectr: OK Null: OK
<SET PARAMETER MODE>
Enter password
v2.41 08-12-09 DM
<SET PARAMETER MODE>
S Factor
1.00000
Enter a value