Troubleshooting
A01-92-4 A .ver 97020-00094
–7
press the * key to enter the number to enter Mode 2 (Set Parameter
Mode - Channel A) or Mode 3 (Set Parameter Mode - Channel B).
2. Press the * key to cycle through the screens until the Peak
Tracking parameter appears.
3. Press 2 (RESET) followed by the * key. The peak tracking function
will cycle off then on and reset.
4. Press the mode key # followed by 1 to return to Mode 1 (Normal
Mode).
Instrument Problems
If the instrument does not appear to be hampered by gas leaks, contamination,
excessive sampling gas temperatures an
d pressures, or electrical noise, refer
to Table A–2 before contacting your sales representative for service.
Removing the Cover
If necessary, the analyzer cover can be removed by first unscrewing the ten
cover screws, as shown in Figure B–1, and gently lifting the cover off the
system taking care not to damage the display and keypad cables. Disconnect
the display and keypad cables if necessary to gain access to the internal
components.
The display and keypad cables are attached to the electronics
inside the housing. Care must be taken when removing the cover
to avoid damaging the cables or connectors. If disconnecting the
cables is necessary, be sure to reconnect each one with the proper
orientation. When reattaching the cover make sure the cables are
not pinched between the body and the cover.
<SET PARAMETER MODE>
S Factor
1.00000
Enter a value
<SET PARAMETER MODE>
Peak Tracking
1
0:O 1:Track 2:Reset