Mode 1: (Normal Mode)
Continuously displays updated measurements. Press the # key followed by the
1 key.
The measurements displayed are:
• H2O — Concentration in units selected in Mode 2.
• P — Pressure in the sample cell in units selected in Mode 2.
• T — Temperature in the sample cell in units selected in Mode 2.
Mode 2: (Set Parameter Mode)
Enables user to view and change measurement parameters. Follow the
procedure under “To change parameters in Mode 2” on page 3-6 for
viewing and changing any of the parameters.
Mode 3: (Not Used)
Mode 4: (System Di
agnostic Parameters)
Displays system diagnostic data for Channel A. These values may be useful
when troubleshooting the system. Press the # key followed by the 4 key.
The diagnostic parameters in Mode 4 displayed are:
• PP2F: Shows the value of the concentration signal in A/D counts. A
normal range is 0 to 8190 depending on the concentration.
+
# 1
<NORMAL MODE>
H2O: 1.0256lb/mmscf
P: 954.4mb T: 76.1F
+
# 4
PP2F:8190 PkDf:34.00
Powr:2538 PkD1:34.00
Indx: 301