SS1000 Operator’s Manual
3 01-92-4 A .ver 97020-0009421
User Sample ID
The User Sample ID parameter assigns a numeric identification to the data
enabling SCADA or DCS systems to associate the data with a particular sample
point.
User Component ID
The User Component ID parameter assigns a numeric identification to the
data enabling SCADA or DCS systems to associate the data with a particular
analyte.
Adjusting Analyzer Reading to Match Specific
Standard(s)
In some instances, the user may wish to adjust the analyzer reading to match
the concentration (or concentrations) of a spec
ific standard (or standards). The
S Factor and S Factor Offset parameters are used to adjust the analyzer
output in the field without affecting the factory calibration. Both parameters are
used when samples from two different concentration standards are available,
whereas only the S Factor Offset parameter is used when a sample from only
one concentration standard is available.
The value of the S Factor parameter,
S, is determined by
,
where
C
1
is the certified concentration of standard No. 1, C
2
is the certified
concentration of standard No. 2,
A
1
is the measured concentration (analyzer
reading) of standard No. 1, and
A
2
is the measured concentration (analyzer
reading) of standard No. 2.
The S Factor Offset parameter,
O, is determined by
,
<SET PARAMETER MODE>
User Sample ID
0
Enter a value
<SET PARAMETER MODE>
User Component ID
0
Enter a value
S
C
2
C
1
A
2
A
1
--------------------
=
O C
1
S A
1
( )=