To check the PkDf and PkD1 values:
For Channel A, press the # key followed by the 4 key to enter Mode 4 (System
Diagnostic Parameters - Channel A) or press the # key followed by the 5 key
to enter Mode 5 (System Diagnostic Parameters - Channel B). A brief wait
screen will appear as the diagnostic data for Channel B is polled.
18. Compare the values for PkDf and PkD1. If the difference is more
than 4, reset the peak tracking function.
To reset the Peak Tracking Function:
1. Press the # key followed by the 2 key (Channel A) or the # key
followed by the 3 key (Channel B). The LCD prompts for a numeric
password. Enter the user password (3142) on the keypad, then
PP2F:8190 PkDf:34.00
Powr:2538 PkD1:34.00
Indx: 301
Zero: -24
<Diagnostic for CHB>
Please wait...
PP2F:8190 PkDf:47.00
Powr:2028 PkD1:47.00
Indx: 253 T: 50.7
Zero: -16 P: 955.7
<SET PARAMETER MODE>
Enter password
v2.40 08-12-08 DM