SS1000 Operator’s Manual
3
01-92-4 A .ver 97020-000948
Measurement and Control Parameters Defined
S Factor
The S Factor parameter is a user definable value that enables adjustment
(without affecting the factory calibration) of the analyzer response (or slope)
in the field [see “Adjusting Analyzer Reading to Match Specific
Standard(s)” on page 3-12].
S Factor Offset
The S Factor Offset parameter is a user definable value that enables
adjustment (without affecting the factory calibration) of the analyzer offset in
the field [see “Adjusting Analyzer Reading t
o Match Specific
Standard(s)” on page 3-12].
# Spectrum Average
The # Spectrum Average parameter sets the number of scans that the
analyzer averages for when determining the concentration. Averaging over
multiple scans lowers noise in the measurement but inherently increases the
response time. Each scan adds about 0.25 seconds to the response time. For
example, if #Spectrum Average is set to “4,an updated concentration value
will be calculated about once every second.
<SET PARAMETER MODE>
S Factor
1.00000
Enter a value
<SET PARAMETER MODE>
S Factor Oset
0.00000
Enter a value (%)
<SET PARAMETER MODE>
# Spectrum Average
16
Enter a value