The data string is comma and space delimited with a carriage return and
includes:
• Concentration - analyte concentration (user selected units)
• Pressure - cell pressure (user selected units)
• Temperature - cell temperature (user selected units)
• PP2F - magnitude of concentration signal (counts)
• Powr - DC signal at absorption peak (counts)
• Indx - position of absorption peak in scan
• Reference Peak PP2F (only for DP systems)
• Reference Peak Powr (only for DP systems)
• Reference Peak Indx (only for DP systems)
• Zero - detected signal with laser turned off (counts)
• PkD1 - present midpoint value (mA)
To capture and save data from the serial port:
1. To save the data from the serial port, use the Transfer/Capture
Text function and enter the Filename to where you would like to
store the captured data.
2. To stop the capture of the serial data, click on Transfer/Capture
Text/Stop.
To read diagnostic data with HyperTerminal:
1. Before entering Mode 6, make sure the serial port on the computer
used for serial communication is connected to the analyzer
and the
output stream is showing on the screen as described in “To launch
HyperTerminal” on page 4-1.
2. To save the data from the serial port, use the Transfer/Capture
Text function and enter the Filename to where you would like to
store the captured data.
The number of seconds between each line of data output should
be the # Spectrum Average number set in Mode 2 or Mode 3
divided by 4.