SS1000 Operator’s Manual
3
– 01-92-4 A .ver 97020-0009441
3. Press the 0 key followed by the * key to turn peak tracking off. Press
the # key followed by the 1 key to return to Mode 1.
4. Press the # key followed by the 2 key. The LCD prompts for a
numeric password.
5. Enter the user password (7758) on the keypad, then press the * key
to enter the number. This puts you into the factory mode.
6. Enter the Phase value (as provided with the calibration documents)
for the desired range followed by the * key.
7. Enter the Midpoint value (as provided with the calibration
document
s) for the desired range followed by the * key.
8. Repeat for Ramp Amplitude, Imod, RCalb, Zero Offset, Span,
PA1, PA2, PA3 and PA4.
9. Press the # key followed by the 1 key to return to Mode 1.
10. Follow the procedure under “To change parameters in Mode 2”
on page 3-6 to enter new parameter values.
When changing parameters in factory mode, be careful not to
change any parameters other than those specified for range
change or the calibration and/or operation of the analyzer may be
adver
sely affected.
<SET PARAMETER MODE>
Enter password
v2.40 08-12-08 DM
<SET PARAMETER MODE>
Phase
320
Enter a value
<SET PARAMETER MODE>
Midpoint
59.70000
Enter a value