14.2 Changing the cartridges
The cartridges have to be changed at regular intervals (12 months) or if the limit value you have
specified is exceeded for a long time. The start of the interval is the point in time at which the
respective filter cartridge was first connected to the system. Proceed as follows to replace the
pretreatment and ultrapure water cartridge:
1. Turn off the OmniaTap xs system by disconnecting it from the mains voltage.
2. Swivel the Optifill dispenser with dispenser arm to the left or right to the side.
3. Remove the cover from the system. To do this, press the two snap locks on the side (see system
description) and pull the cover hood forwards off the main housing.
4. The ultrapure water cartridge is on the left parking space, the pre-treatment cartridge is on
the right parking space (see chapter assembly). Open the quick-release fasteners by pressing
the clip (see figure) on the system's quick-release fastener and remove the cartridge to be
changed.
5. Place the new cartridge in the appropriate slot in the system. Use the positioning aid on the
bottom of the housing for easier positioning.
6. Connect the cartridge to the system using the quick-release fasteners. Make sure that the
quick-release fasteners engage audibly.
7. Enter the code printed on the filter cartridge in the customer menu under the item "Change
filter set" and confirm your entry.
8. Replace the cover and turn the OptiFill dispenser back to the desired position.
9. Plug the power cord back in to start the system.