OmniaTap xs
basic
; Version August 2023
Subject to technical changes
Depending on the country, a safety combination consisting of a backflow preventer and a system
separator may have to be available on the building side.
A DN 50 floor drain with free drainage is to be provided in the operating room. If there is no floor drain,
we recommend using a water monitor to prevent damage to the system and equipment as well as the
building. Otherwise, the manufacturer assumes no liability for any water damage that may occur.
The base of the system must have sufficient load-bearing capacity (for weight, see technical data).
When assembling, make sure that there is enough free space for problem-free operation, maintenance
and repairs.
Maintenance work may only be carried out by trained specialists.
3 Intended Use
The systems of the OmniaTap xs series are used for the direct treatment of drinking water into pure
and ultrapure water for laboratory use. In order to ensure maximum quality of the pure and ultrapure
water with the longest possible service life of the consumables, the OmniaTap xs system must be fed
with drinking water in accordance with DIN 2000. Any other use of the system is not permitted and is
considered improper use.
The pure / ultrapure water produced is used as a solvent or rinsing agent in a wide variety of analysis
methods such as high-performance liquid chromatography (HPLC), ion chromatography (IC), atomic
absorption spectrometry (AAS), ultra-trace analysis, etc. It is also used in a variety of chemical and
biochemical applications such as preparing reagents, cultivating cells, etc.
The treated water is not suitable for consumption.
The system must not be used for the production of pharmaceuticals and is not a medical device.