19.2 Faults table
Display does not function
No or faulty power supply
Ensure the power supply
according to the specifications
under "6 Technical data".
Conductivity LF1 permanently
exceeds the set limit value /
error LF1!
exhausted
cartridge
Feed water does not meet the
requirements
Check the quality of the feed
water
Limit for LF1 set too low
LF1
Conductivity LF1 is displayed
incorrectly and the error LF1!
Is issued
Defective conductivity
measurement LF1
Please contact customer
service
Conductivity LF2 permanently
exceeds the set limit value /
error LF2!
Ultrapure water cartridge is
exhausted
Replace the ultrapure water
cartridge
exhausted
cartridge
Feed water does not meet the
requirements
Check the quality of the feed
water
Limit for LF2 set too low
Check the setting for limit for
LF2
Conductivity LF2 is displayed
incorrectly and the error LF2!
Is issued
Defective conductivity
measurement LF2
Please contact customer
service
Temperature permanently
exceeds the set limit value /
error TEMP!
Temperature limit set too low
Check the temperature limit
setting and the temperature of
the feed water
The system is set to permanent
circulation or the pause
between the circulation phases
is too small
Increase the value of the Pause
parameter in the circulation
settings
UV lamp not working / error
UV!
UV lamp or UV ballast
defective
UV lamp or UV ballast needs to
be replaced.
Maximum operating time has
been exceeded
System does not produce pure
water, tank is not filled
System is not receiving feed
water
Check the hose connection of
the feed water supply. Open
the domestic feed water
supply.