Monitor Setup
12 M200 Fall Monitor – Setup & User Guide
Monitor Setup
If the optional closed-window cover is being used, loosen the battery
cover screw and remove the battery cover to reveal the control settings.
Volume (Speaker icon)
There are 4 Volume settings for the Pad Alarms:
• 1 - Low
• 2 - Medium
• 3 - High
• NC (Nurse Call) - If a Nurse Call jack is plugged in and this setting is
selected, the local alarm sound is suppressed completely, and the
alarm is sent to the Nurse Call station. If NC is selected and there is
no Nurse Call cable connected, the volume will be high.
Tone Selection (Music Note icon)
There are 4 different Alert tones available for selection:
• Low-pitched series of 3 beeps followed by 2 beeps
• High-pitched fast series of Low and High beeps
• Low-pitched slow series of High and Low beeps
• Low-pitched fast series of Low and High beeps