Monitor Setup
16 M200 Fall Monitor – Setup & User Guide
Latching and Non-Latching
Monitor
Latching: Requires staff to clear all local alarms by pressing the Reset
Button.
Non-Latching: Allows the monitor to be reset by re-applying weight to
the pad or by pressing the Reset Button.
Nurse Call Output
Note: The following descriptions apply to the Nurse Call output of the
monitor. Your Nurse Call system may behave differently depending
on how your system is configured. Test before using with the
Nurse Call system. See “Interfacing with a Nurse Call System” on
page 30.
Latching: Nurse Call output stays active until the staff has cleared the
monitor by pressing the Reset Button.
Non-Latching: Nurse Call output stays active until either weight has
been reapplied to the pad or the Reset Button has been pressed on the
monitor.
Note: Floor Pads always follow Latching behavior.