Cleaning
22 M200 Fall Monitor – Setup & User Guide
Floor Pad
1. Connect the Floor Pad to the Monitor. If two pads are required,
connect both pads and test each pad individually.
2. Step onto the Floor Pad.
3. Verify the Monitor generates a Pad Alarm. The Alarm State
Indicator and Reset Button should continuously flash red and sound
a local audible alert.
Note: If a Nurse Call system is connected and the volume is set to NC, there is
no local audible alert.
4. If the Monitor is connected to a Nurse Call system, verify that an
alarm was triggered on that system.
5. Press the Reset Button on the Monitor to cancel the alarm.
6. If a Low Battery message is announced at this time, replace all
batteries (see “Replacing Monitor Batteries” on page 25) and
repeat the above test procedure.
7. If the Monitor Pad Alarm cannot be verified as above, check all
connections and repeat this test procedure.
8. If the Pad Alarm still does not sound, DO NOT place the pad or
Monitor into service and follow your facility’s procedures for
product maintenance.
Cleaning
The monitor and the Sensormat pad may be cleaned with a damp cloth
or sponge using mild disinfectants. Never use alcohol, acidic or harsh
petroleum-based cleaners.
Note: The Sensormat Pad is for single patient/resident use only.