Testing the Sensormat Pads
20 M200 Fall Monitor – Setup & User Guide
Testing the Sensormat Pads
Test the Sensormat Pad and M200 monitor before first use, each time the
system is put into use, and daily thereafter.
Bed, Chair, Toilet and Commode
1. Connect the pad to the monitor. If two pads are required, connect
both pads and test each pad individually.
2. Apply FULL and FIRM palm pressure with your hand to the
Sensormat pad and maintain pressure for at least three seconds
while the Monitor arms.
3. Verify the Reset Button LED flashes green during the three second
period, and is followed by a Positive Confirmation Tone.
4. Release hand pressure.
5. After a delay of up to two seconds (depending on configured
delay settings), verify the Monitor generates a Pad Alarm. The
Alarm State Indicator and Reset Button should continuously flash
red and sound a local audible alert.
Note: If a Nurse Call system is connected and the volume is set to NC, there
is no local audible alert.
6. If the Monitor is connected to a Nurse Call system, verify that an
alarm was triggered on that system.
7. Press the Reset Button on the Monitor to cancel the alarm.
8. If a Low Battery message is announced at this time, replace all
batteries (see “Replacing Monitor Batteries” on page 25) and
repeat the above test procedure.
9. If the Monitor Pad Alarm cannot be verified as above, check all
connections and repeat this test procedure.
10. If the Pad Alarm still does not sound, DO NOT place the pad or
Monitor into service and follow your facility’s procedures for
product maintenance.