4-8
P129391-130 Operator/Maintenance Manual Installation
Adjust Travel Stops (Double Door)
If sterilizer is a double door unit adjust the front caster travel stops as
follows:
STEP 1: Push transfer carriage with empty loading car into sterilizer
chamber
• Stop when center line of front rollers are about 7/8"
(22 mm) from far end of sterilizer chamber track.
STEP 2: Adjust rear stops on transfer carriage to contact front
caster leg carrier.
• Lower lift mechanism on transfer carriage until the
loading car rests on sterilizer tracks.
STEP
1
Loading Car
Sterilizer Chamber
Push loading car into chamber until front
rollers are 7/8" (22 mm) from end of track
STEP
2
Adjust Rear Stop To
Contact Top of Rear
Leg