Note
If an electrolyte with a significantly different viscosity than water is used or the
equipment is operated at sub-zero temperatures, you may have to adjust the flow
using the viscous electrolyte or at low temperature.
This function calibrates the pump of the polishing unit and ensures that the flow rate settings in the
Struers methods are correct.
Examine the state of the electrolyte before you adjust the pump.
Use this function to check the pump regularly.
Use this function to calibrate the pump only if the correct results are no longer achieved.
Procedure
Carry out this procedure from the control unit.
1. From the Main menu screen, select the Configuration screen.
2. Press F4 - Adj. Pump.
3. Fill an electrolyte container to the max. mark with water.
4. Add one drop of detergent to release the surface tension of the water.
5. The following message is shown:
Insert container
Please insert a container filled to the max. mark with water.
Add a drop of detergent.
6. Insert the container in the polishing unit.
7. Press Enter to continue.
8. Remove the mask and insert the tube supplied with the machine into
the outlet for the electrolyte.
9. Press Enter to continue.
10. Select Maximum pump flow.
11. Press Enter to start the pump.
12. Adjust the flow so that the water reaches the maximum mark on the
tube.
13. Press Enter to save the value.
14. Select Minimum pump flow.
26 LectroPol-5