The equipment is intended for use in quality control applications, where the surface can be prepared
for further materialographic inspection with a transmitted electron microscope (TEM).
The equipment is designed for preparation of conductive materials suitable for electrolytic etching.
In order for the equipment to operate correctly and safety, it must be used with Struers accessories
and consumables specially designed for this purpose and this type of device.
When in use, the machine equipment must not be touched, moved or tampered with.
The operator must be fully instructed in how to handle and use electrolytes with this machine.
The machine must be operated only by skilled/trained personnel.
The machine must be connected to a cooling unit in order to function as intended.
The polishing unit must be placed on an even surface.
The polishing unit must be placed in a well ventilated area, preferably a fume cabinet.
Control unit
See the Instruction Manual for this unit.
Do not use the machine for
the following
Preparation of materials other than materials suitable for
materialographic studies.
The machine must not be used for any type of explosive and/or
flammable material, or materials which are not stable during
machining, heating or pressure.
Do not use the machine without sufficient ventilation.
The machine must not be used with consumables or a
combination of electrolytes and accessories which are not
compatible for use with this equipment.
Model LectroPol-5
2.2 LectroPol-5safety precautions
2.2.1 Read carefully before use
1. Ignoring this information and mishandling of the equipment can lead to severe bodily injuries
and material damage.
2. The machine must be installed in compliance with local safety regulations. All functions on the
machine and any connected equipment must be in working order. The machine must be earthed
(grounded).
3. The operator must read the safety precautions and Instruction Manual, as well as relevant
sections of the manuals for any connected equipment and accessories.
4. The operator must be fully instructed in how to handle and use electrolytes with this machine.
5. Follow all safety requirements for handling, mixing, emptying and disposing of electrolytes.
6 LectroPol-5
2 Safety