3. Press Enter to continue.
4. Follow the on-screen instructions.The on-screen instructions are listed below.
5. Press Enter to continue through the sequence of steps.
You can cancel the process at any time. To do so, press Back.
6. The following messages appear. Follow each step listed on the screen.
[Remove Electrolyte:]
1. Lift the polishing table.
2. Please wait for 59 s
3. Remove the present elec.
[Insert Water]
1. Insert a container with water.
2. Lower the polishing table.
[Cleaning...]
The system is being cleaned now.
Please wait 57s
[Cleaning done]
1. Lift the polishing table.
2. Use alcohol to remove water.
3. Clean off the remaining alcohol.
[Remove water]
Remove the water.
7.3 Monthly
7.3.1 Calibrating the pump
If the results are not correct, or if you cannot reproduce the results, adjust the pump.
Note
The pump must be adjusted before it is used for the first time.
7 Maintenance and service
LectroPol-5 25