3. Open the cover of the polishing unit.
The power to the polishing unit must be interrupted immediately. This means
that the polishing current must drop to 0.
4. If operation does not stop, press the Stop button.
5. Contact Struers Service.
With the protective cover open
1. Start a process: Press the Start button.
The machine starts operating.
The pump begins to circulate electrolyte and the electrical connection is
checked.
The power to the polishing unit must be interrupted immediately. This means that
the polishing current must drop to 0. The Current-time graph must show that the
polishing current is 0.
2. An error message appears: No electrical connection. Check anode arm.
3. If the polishing current is not 0 and electrolytic polishing begins, press the Stop
button.
4. Contact Struers Service.
7.5 Spare parts
For specific safety related parts, see the section "Safety Related Parts of the Control System
(SRP/CS)" in the section "Technical data" in this manual.
Technical questions and spare parts
If you have technical questions or when you order spare parts, state serial number and
voltage/frequency. The serial number and the voltage are stated on the type plate of the machine.
For further information, or to check the availability of spare parts, contact Struers Service. Contact
information is available on Struers.com.
7.6 Service and repair
We recommend that a regular service check be carried out yearly or after every 1500 hours of use.
Note
Service must only be performed by a qualified technician (electromechanical,
electronic, mechanical, pneumatic, etc.).
Contact Struers Service.
Service check
We offer a range of comprehensive maintenance plans to suit the requirements of our customers.
This range of services is called ServiceGuard.
28 LectroPol-5
7 Maintenance and service