9.3 Safety Circuit Categories/Performance Level
Safety Circuit Categories/Performance Level
Work zone interlock EN 60204-1, Stop category 0
EN ISO 13849-1, Category 1
PLc
9.4 Safety Related Parts of the Control System (SRP/CS)
WARNING
Safety critical components must be replaced after a maximum lifetime of 20 years.
Contact Struers Service.
Note
SRP/CS (safety-related parts of a control system) are parts that have an influence
on safe operation of the machine.
Note
Replacement of safety critical components must only be performed by a Struers
engineer or a qualified technician (electromechanical, electronic, mechanical,
pneumatic, etc.).
Safety critical components must only be replaced by components with at least the
same safety level.
Contact Struers Service.
Safety
related
part
Manufacturer/Manufacturer
description
Manufacturer
catalog no.
Electrical
ref.
Struers
catalog no.
Interlock
switch
circuit -
Polishing
unit
Schmersal
Coded-magnet sensor
BNS33-11Z-2M SS1 2SS00140
Interlock
switch
circuit -
Polishing
unit
Schmersal
Coded-magnet (actuator) for
sensor
BPS33 SS1 2SS00141
Interlock
switch
circuit -
Polishing
unit
Finder
Power relay
62.32.9.024.4800 K1 2KL46680
9 Technical data
LectroPol-5 31