EN-21
Console Controls
Adjusting CUT Power
• Press the adjust
buttons on the console
• Press the gray button on the handpiece (increase)
• Press the I (decrease) and III (increase) buttons on the footswitch
Adjusting COAG Power
1. Press and hold menu
. The COAG POWER LEVEL
screen will appear.
2. Press adjust
to adjust.
3. Press select
to confirm selection and exit.
Note: COAG power can only be adjusted when an RF probe
is connected to the console.
COAG1
COAG2
COAG3
COAG POWER LEVEL
Selecting Force Modulation
The Crossre2 Console features an additional RF mode known as Force Modulation. Force
Modulation is an alternative ablation mode that duty cycles RF output at a low frequency to
achieve a lower average power output than in normal CUT mode.
Currently, Force Modulation is an option only with the following SERFAS Energy probes: 90-S, 90-S
Max, and Super 90-S.
• To activate Force Modulation, hold down the grey power button on the SERFAS probe for
three seconds. A hammer icon will appear on the LCD screen of the console, indicating
Force Modulation activated.
• To deactivate Force Modulation, hold down the grey power button on the SERFAS probe for
three seconds. The hammer icon will disappear from the LCD screen.

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals