EN-11
Setup and Device Connections
Stryker Endoscopy considers instructional training an integral part of the Crossre2 system. Your
Stryker Endoscopy sales representative will perform at least one inservice at your convenience
to help you set up your equipment and instruct you and your sta on its operation and
maintenance. Please contact your local Stryker Endoscopy representative to schedule an in-
service after your equipment has arrived.
Warning
• Be sure that no liquid is present between connections to the console and the
handpiece. Connection of wet accessories may lead to electric shock or electrical
short.
• To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
• Use only hospital-grade power cables. Using other cables may result in increased RF
emissions or decreased immunity from such emissions.
• Only the handpieces and disposable attachments are suitable for use in the patient
environment. The console and footswitch are not sterile devices and should not enter
the sterile eld.
• The Crossre2 System is compatible only with the Stryker handpieces and
footswitches listed in this manual. Do not connect any equipment not specied in
this manual, as unexpected results or serious injury will occur.
• The separable AC power cord is provided as a means of emergency shutdown and
disconnection from the power source. Do not position the console in a way that is
dicult to disconnect the AC power cord.
1. Place the console on a sturdy platform, such as a Stryker cart.
• Select a location according to the recommendations in the “Electromagnetic
Compatibility” section of this user guide.
• Leave four inches of space around all sides for convection cooling.
• Do not obstruct the fans located near the rear and side of the console. Position the
console so the fan directs the ow of air away from the patient.
• Keep the activation indication lights and speaker in eld of view and hearing at all
times during activation. The light and sound are important safety features.

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals