EN-3
Warnings and Cautions
Caution: Federal law (USA) restricts this device to use by, or on order of, a physician.
Operator Prole
The Crossre2 system is intended for use only by licensed medical professionals, properly
trained in the use of arthroscopic and electrosurgical equipment and techniques. The Crossre2
system generates potentially hazardous levels of energy that can result in injury or even death if
improperly used.
General Warnings
To avoid potential serious injury to the user and the patient, observe the following warnings:
1. Read this manual thoroughly and be familiar with its contents prior to operating the
equipment.
2. Carefully unpack the device and ensure that all components are accounted for and remain
undamaged from shipment.
3. Inspect all handpieces and probes for damage to the cable insulation. If damage is found,
refer to the Stryker Standard Warranty and Return Policy (1000-401-175).
4. Before using the Crossre2 system in an actual procedure, verify that each component is
installed and functioning properly. Improper connection may cause arcing or malfunction of
the handpiece or console, which can result in injury, unintended surgical eect, or product
damage.
5. Do not use the Crossre2 system on patients with cardiac pacemakers or other electronic
device implants. Doing so could lead to electromagnetic interference and possible death.
6. Do not attempt to reuse or resterilize any product labeled “Single-Use, as this may lead to
equipment malfunction, patient/user injury, and/or cross-contamination.
7. Shaver handpieces are provided nonsterile and must be cleaned and sterilized prior to each
use, according to the reprocessing instructions provided in the handpiece manual.
8. Do not use the Crossre2 system with non-conductive irrigants (e.g. sterile water, air, gas,
glycine, etc.). Use only conductive irrigants such as saline or Ringers lactate in order for the
system to function properly.
9. Do not activate the Crossre2 system for prolonged lengths of time when the attachment is
not in contact with tissue. Doing so may lead to unintentional damage to surrounding tissue.
10. Do not obstruct the fans located near the rear and side of the console. Position the console
so the fan directs the ow of air away from the patient.
11. Keep the activation indication lights and speaker in eld of view and hearing at all times
during activation. The light and sound are important safety features.

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals