EN-35
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The Crossre2 System is intended for use in the electromagnetic environment specied below. The customer or the user of
Crossre2 System should ensure that it is used in such an environment
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment - guidance
Electrostatic Discharge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact ±8kV air Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast transient/burst
IEC61000-4-4
±2kV for power
supply lines
±2kV for power supply lines Mains power quality
should be that of a typical
commercial or hospital
environment
±1kV for input/
output lines
±1kV for input/output lines
Surge IEC61000-4-5 ±1kV dierential
mode
±1kV dierential mode Mains power quality
should be that of a typical
commercial or hospital
environment
±2kV common mode ±2kV common mode
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC61000-4-11
<5% Ut (>95% dip in
Ut) for 0.5 cycle
<5% Ut (>95% dip in Ut) for 0.5 cycle Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of Crossre
2 requires continued
operation during power
mains interruptions, it
is recommended that
Crossre 2 be powered
from an uninterruptible
power supply or a battery.
40% Ut (60% dip in
Ut) for 5 cycles
40% Ut (60% dip in Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
70% Ut (30% dip in Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec
<5% Ut (>95% dip in Ut) for 5 sec
Power frequency (50/60Hz)
magnetic eld IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration--Electromagnetic Immunity
Crossre2 System is intended for use in the electromagnetic environment specied below. The customer or the user of
Crossre2 System should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment--Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Crossre 2
system, including its cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
Recommended Separation Distance:
d = 1.2√P

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals