EN-5
4. When the Crossre2 system and physiological monitoring equipment are used
simultaneously on a patient, position any monitoring electrodes as far as possible from the
surgical electrodes. Monitoring equipment using high frequency, current-limiting devices is
recommended. Needle monitoring electrodes are not recommended.
5. During use, operators should wear standard surgical gloves to help reduce the risk of electric
shock.
6. To prevent patient injury, select the lowest output power required for the intended purpose.
7. Do not exceed the rated accessory voltage of electrosurgical accessories. Only use
electrosurgical accessories that have a rated accessory voltage equal to or greater than the
maximum output voltage of the generator.
8. Do not activate the Crossre2 system until the probe is properly positioned in the patient.
9. Ensure that the probe tip, including the return electrode, is completely surrounded by
irrigant solution during use.
10. Maintain the active electrode in the eld of view at all times to avoid tissue damage.
11. Keep active electrodes isolated from the patient when not in use.
12. When not in use, remove the handpiece and disposable attachments from the surgical
site and place them away from metallic objects. Attachments should be separated from
other electrosurgical equipment to avoid inadvertent electrical coupling between devices.
Inadvertent activation may cause user/patient injury and/or product damage.
13. Failure of the system may result in an unintended increase in output power.
14. Neuromuscular stimulation may occur when RF probes are used.
15. Smoke generated during electrosurgical procedures may be harmful to surgical personnel.
Take appropriate precautions by wearing surgical masks or other means of protection.
Cautions
To avoid product damage, observe the following cautions.
1. While using the handpiece, do not touch the attachment to metal objects, such as an
endoscope or metal cannula. Damage to the attachments or other devices may result.
2. Attempt no internal repairs or adjustments, unless specied otherwise in this manual. Units
requiring repair should be returned to Stryker.
3. Pay close attention to the care and cleaning instructions in this manual. Failure to follow
these instructions may result in product damage.
4. Do not remove the cover of the console as this could cause electric shock and product
damage.

Other manuals for Stryker Crossfire 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Crossfire 2 and is the answer not in the manual?

Stryker Crossfire 2 Specifications

General IconGeneral
BrandStryker
ModelCrossfire 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals