Page #22 background imageLoading...
Page #22 background image
Service Information − Optional Scale System
21
SCALE DISPLAY BOX REPLACEMENT
Required Tools:
Wire Cutters 1/2” Open End Wrench Rivet Gun T27 Torx
1. Pump the litter up to full height.
2. Using a T27 Torx, remove the screws holding the litter foot end cover and the foot end corner covers.
3. Using wire cutters, cut all cable ties and disconnect the cables from the load cells.
4. Using a 1/2” socket, remove the six bolts holding the litter foot end weldment. Save the bolts for reuse.
5. Drill out the rivets holding the scale display box to the foot end weldment.
6. Plug the load cell cables into the display cables to verify the system is working, then unplug the cables.
7. Using two rivets, attach the new scale display box to the foot end weldment.
8. Using the six bolts removed in step 3, reattach the foot end weldment to the siderail tubes.
9. Reinstall the corner covers and the foot end cover,
10. Route the display cables and connect them to the load cell cables.
11. Using a known weight, verify scale function with the litter flat and in approximately 10 degrees Trend and
reverse Trend.
Return to Table of Contents

Table of Contents

Other manuals for Stryker M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker M Series and is the answer not in the manual?

Stryker M Series Specifications

General IconGeneral
BrandStryker
ModelM Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals