Control Panel and Display 10
Power Options
AC Power
The Smart Pump’s primary power source is AC power. As a safety
feature, the SmartPump’s internal battery provides an alterative (back
up) power source if AC power is lost.
Always connect the SmartPump to an AC power source for normal
operation.
Internal Battery
The SmartPump’s internal battery automatically provides back-up
power if AC power is interrupted. The internal battery when fully
charged supports up to five hours of operation.
If, at initial start-up, the SmartPump is not plugged into AC power, the
following occurs:
♦ no AC is displayed on the lower left of the LCD.
♦ The Alarm Indicator Mute button flashes red.
Pressing the Alarm Indicator Mute button within 30 seconds of initial
start-up will cancel the alarm. In this mode, the Alarm Indicator Mute
button will return to green, the no AC indicator will remain illuminated
and the battery symbol will indicate the amount of battery charge.
NOTE: If, at initial start-up, there is no AC present and the Alarm
Indicator Mute button is not pressed within 30 seconds, the
SmartPump will return to its ‘OFF(standby)’ mode automatically. This
feature prevents the SmartPump from unintentionally being used in a
(back up) battery mode and prevents accidental battery discharge
during transport and storage.
If AC power is interrupted during normal use, the SmartPump will use
its internal battery automatically. Press the Alarm Indicator Mute
button to acknowledge the change in power source. The Alarm
Indicator Mute button returns to green.
When AC power is restored, the SmartPump will return to its normal
AC operation automatically.
When the SmartPump is unplugged and turned ‘OFF(standby),’ it
conserves its internal battery automatically using a power conservation
mode. If fully charged, the internal battery is kept charged for up to
750 hours.