Control Panel and Display 13
Control Buttons
Button Explanation
♦ Turn ON by pressing, then releasing the ON/OFF (standby)
button.
♦ Turn OFF by pressing and holding the ON/OFF (standby)
button for 1.5 seconds (safety pause).
WARNING: Turning OFF the SmartPump while cuffs are inflated
will deflate the cuffs and total elapsed time information will be
lost.
SET/SAVE
SET/SAVE
1. Press the Set/Save button to initiate a change of the
TIME
and/or target
PRESSURE for the corresponding cuff. The display will
blink after the Set/Save button is pressed.
2. Press the Set/Save button again to save new time or pressure
values.
+
TIME
-
T
IME Increase/Decrease (located above the Set/Save button)
After pressing the Set/Save button, press these buttons to
increase (+) or decrease (-) the inflation time for the selected
cuff. Each button press changes the time value in 5-minute
increments (1-minute increments within the 1 to 15 minute time
range). The numeric display changes accordingly.
+
PRESSURE
-
P
RESSURE Increase/Decrease (located below the Set/Save button)
After pressing Set/Save button, press these buttons to increase
(+) or decrease (-) the inflation pressure for the selected cuff.
Each button press changes the pressure value in 5 mmHg
increments. The numeric display changes accordingly.
INFLATE
I
NFLATE
Press the
INFLATE button of the corresponding cuff to inflate the
cuff to its target pressure. If a cuff is deflated during a procedure,
press the Inflate button to return to the set pressure. The Timer
stops during the period of a deflation and resumes accumulating
total ‘cuff time’ upon reinflation.