Adding external attenuators. If EMI becomes an unusually difficult problem, external devices such as an
isolation transformer or a transient suppressor may be of help. Your Service Provider can be of help in
determining the need for external devices.
Where it has been established that electromagnetic interference is affecting physiological parameter
measurement values, a physician, or a suitably qualified person authorized by a physician, should determine
if it will negatively impact patient diagnosis or treatment.
Electromagnetic Emissions and Immunity
The monitor is suitable for use in the electromagnetic environment specified in the table below. You must
ensure that it is used in such an environment.
Table 1 - Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The SRF618B6 is intended for use in the electromagnetic environment specified below. The customer or the user
of the SRF618B6 should assure that it is used in such an environment.
Emissions Compliance Electromagnetic environment-- guidance
RF emissions
CISPR 11
Group 1
The SRF618B6 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The SRF618B6 is suitable for use in all establishments, but if
used in domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes, whatever additional
measures are necessary.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity
The monitor is suitable for use in the specified electromagnetic environment. The user must ensure that it is
used in the appropriate environment as described below.
Table 2-Guidance and manufacture’s declaration – electromagnetic immunity
The SRF618B6 is intended for use in the electromagnetic environment specified below. The customer or the user of
the SRF618B6 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
--guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
± 2 kV line(s) to earth
±1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of
a typical commercial or hospital