~47~
Chapter 9 After Monitoring
9.1 Completing Monitoring
After monitoring,
1) Remove transducers or electrodes from the patient; wipe the remaining gel off the patient and the
transducer with a clean soft cloth or tissue.
2) Long press the PRINT key to stop printing.
3) Wait the paper to stop and then tear it off along the perforation.
NOTE:
After the fetus is delivered in the labor, the monitor may pick up signals of the umbilical cord and display a
trace/numeric. To avoid misinterpretation, it is recommended to remove the transducers from the patient and
switch off the monitor immediately after the fetus is delivered.
9.2 Switching Off
1) Press and hold the POWER key for at least 3 seconds to switch off the monitor.
2) Unplug the power cord.
CAUTION:
Do not press the POWER key
continuously. Allow at least 10 seconds between switching the monitor
on and off.