EasyManuals Logo

Supco VLTS User Manual

Supco VLTS
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Temperature recorder and printer
19
8. After selecting special DC sensors, such as 0-10V, 0-5V, 0-20mA, 4-20mA, which measure
low voltage or current, an adjustment of measured value limits to displayed limits should be
set by scaling, enter a minimum and a maximum value, respectively, and select one mark
for the units. When Sensor No.1 scaling, max value: appears, select with / to
change the number to be displayed when a maximum value is measured. For confirmation,
press . When Sensor No.1 scaling, min value: appears, select with / to
change the number to be displayed when a minimum value is measured. For confirmation,
press . When Sensor No.1, measurement unit: appears, select with / to
change the unit to be displayed. For confirmation, press , and a confirmation such as
Channel No. 1 set will appear. To select another channel number for configuring a sensor,
press / or to exit.
On/off (dry) Contact opening sensor: connect the white wire (Pin6-Data) to the blue
wire (Pin1+5V).
DC sensors (analog) such as 0-10V, 0-5V, 0-20mA, 4-20mA. Connect
the black wire (Pin1-GND) to the white (Pin6-Data).
Please note: do not use a power supply of voltage higher than 24V.
Analog DC sensors may be connected to the contact sensor cable.
Note: the sensor sockets (RJ11) have 6 pins, as in the diagram below.
The left pin, no. 1 in the socket, is a blue wire in the plug.
The * character in the settings screens indicate an option (not yet selected).
To exit the menu and revert to readings mode, press [x} twice. Please wait, saving
setting will be displayed, then several seconds later, the readings screen. At the same time,
Settings saved will be printed.
Notes: any change in the sensor type will boot the statistics in all the channels, will rename
the channel and change the channels alarm settings to the default value.
SENSOR CALIBRATION / TEMPERATURE CHANNEL
To calibrate the sensors/temperature:
1. Press / until the Settings Submenu is displayed.
4. Press to confirm entry into the settings menu. The display will show Enter
password:
5. Enter a password (4 keys in a row). If you have not yet changed your password, use the
default password: , , , .
6. Go to the Set Calibration menu using the / keys, and press .
7. Select Channel No. 1 will appear, the channel number will flash. To change the channel
number (1-4), press / and confirm with .
8. The current calibration value in the channel will flash, for example Value: 0.0.
9. Press / to change the magnitude of the calibration deviation for the channel, in
0.1 increments. The measurement deviation can be change in the range of from +8 to -8.
10. Press to confirm. Select channel No. 1 will appear again.
11. Press / to calibrate another channel or twice to complete and save the changes.
Please wait, Saving Setting will appear for a few seconds followed by the readings
screen. At the same time, Settings saved will also be printed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Supco VLTS and is the answer not in the manual?

Supco VLTS Specifications

General IconGeneral
BrandSupco
ModelVLTS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals