36
6.4.2 Dynamic Span Calibration
Let the belt weigher run with loading measured materials or Poise Weights for correcting Span
Coefficient.
After running the last revolution:
End 5R
[ENT]: Input
2 Span Cal. Dyn.
W: 499.2kg
ENT
Flow: 12.6t/h
Wait 1.00m
2 Span Cal. Dyn.
W: 489.2kg
Main Display Interface
MENU+】【 : F2 Calibration
ENT+】【 : 2 Span Cal. Dyn.
ENT& Start loading
M
Remainder length of a revolution
Flow: 12.6t/h
01R 10.00m
2 Span Cal. Dyn.
W: 1.0kg
Revolution Count
】【 :
Display Flow, Load , AD Value
Stop loading &ENT
0000500.0[kg]
R:0.1-9999999.9
Input Weight